Selecting suitable mechanical thermometers for a process is not an instant task. Bimetal thermometers or gas-actuated thermometers ? your choice on the sort of thermometer alone depends on several factors. The most important selection criteria is the response time of the measuring Instrument.
When it comes to speed, there is no way around gas-actuated thermometers. Generally of thumb, these thermometers take about 30 seconds for a temperature adjustment from 10�C to 90�C. A bimetal thermometer would display exactly the same final value after 90 seconds. Due to their short response times, gas-actuated thermometers are particularly suitable for integration into switching requirements.
The difference in speed is explained by the different functional principles of both thermometer types ? unlike the bimetal thermometer (which uses the principle of different expansion coefficients), the measuring system of a gas-actuated thermometer is filled up with helium under pressure. Any change in temperature causes an instantaneous change in internal pressure in the stem. The pressure deforms the measuring tube and the deflection is used in the pointer with a movement.
Thermowells affect the response time
As with other measuring instruments, in critical processes, mechanical thermometers must be coupled with a thermowell. It must be considered that this type of sleeve may also have an influence, to some degree significant, on the response time of the thermometer. To what Plummet , depends upon the material and wall thickness of the thermowell.
Note
Further details on mechanical thermometers are available on the WIKA website under the headings Dial thermometers and Thermometers with switch contacts.
See also our articles
Selection criteria for mechanical thermometers (2): process temperature
Selection criteria for dial thermometers (3): Capillary